Chapter 6. Defining Study Setups
6.8 Specifying Calibration Parameters
Figure 36: Calibration Parameters Screen
The Kaye Validator provides both sensor calibration and calibration verification. You specify
all calibration parameters on the Calibration Parameters screen.
Before you perform a qualification study, calibrate the temperature sensors to correct raw
temperature readings to a NIST-traceable temperature standard. During calibration the Kaye
Validator automatically corrects raw temperature sensor readings to a NIST-traceable
temperature standard to compensate for any errors inherent in sensor materials, and stores
these offsets in the associated SIMs. You can specify a one-point calibration, a two-point
calibration, or a two-point calibration and a one-point check. Sensors that do not meet the
calibration test criteria are marked as failed and are not calibrated.
After the qualification study, the Kaye Validator performs a calibration verification to verify
that the sensor readings are still within the calibration criteria. You can specify a one-point,
two-point or three-point calibration verification. If a sensor fails calibration verification, the
sensor is noted as failed but the readings are still reported. Calibration offsets are not changed
during calibration verification.
70
Kaye Validator User’s Manual
Summary of Contents for Validator AVS
Page 1: ...Validation Kaye Validator AVS User s Manual Kaye Validator User s Manual...
Page 2: ...Kaye Validator AVS Thermal Process Validation System User s Manual M5100 EN Rev A May 2015...
Page 3: ...No content intended for this page Kaye Validator User s Manual...
Page 133: ...Chapter 10 Generating Reports 126 Kaye Validator User s Manual...
Page 138: ...Chapter 10 Generating Reports Kaye Validator User s Manual 131...
Page 147: ...Chapter 10 Generating Reports 140 Kaye Validator User s Manual...
Page 167: ...No content intended for this page...