Chapter 2. The Validator Hardware
2.6 Connecting the System
All system connections are on the back of the Validator. Make sure the unit is powered off
before connecting the IRTD, temperature reference (LTR-90, LTR -25/140, LTR -40/140,
HTR 400, CTR -80, or CTR -40), and SIMs
To connect your Validator, complete the following steps:
1.
Connect the IRTD to one of the 4 pin Din sockets, using the cable supplied for IRTD.
2.
Connect the temperature reference to one of the 4 pin Din sockets, using the
cable supplied for temperature reference.
3.
Insert the wired SIMs into their associated slot positions.
4.
Insert the thermocouple sensors into the temperature reference. See the previous section,
Sensor and IRTD Installation
, for installation instructions.
5.
Connect your Console to the docking station, Ethernet or USB port, using the cables
provided.
6.
(Optional) Connect a printer to the Console.
7.
Connect the temperature reference power cable and the Validator power cable to
a grounded power source.
8.
Turn the temperature reference power switch to the ON position.
9.
Turn the Validator power switch to the ON position.
IMPORTANT:
SIMs are only recognized by the system when you power up the Validator.
If a SIM is disconnected after power up, that SIM's sensors will be read as
Open and no data will be recorded for that SIM. To rectify this, power
down and power back up.
34
Kaye Vali
dator User’s Manual
Summary of Contents for Validator AVS
Page 1: ...Validation Kaye Validator AVS User s Manual Kaye Validator User s Manual...
Page 2: ...Kaye Validator AVS Thermal Process Validation System User s Manual M5100 EN Rev A May 2015...
Page 3: ...No content intended for this page Kaye Validator User s Manual...
Page 133: ...Chapter 10 Generating Reports 126 Kaye Validator User s Manual...
Page 138: ...Chapter 10 Generating Reports Kaye Validator User s Manual 131...
Page 147: ...Chapter 10 Generating Reports 140 Kaye Validator User s Manual...
Page 167: ...No content intended for this page...