Appendix A. Environmental Compliance
System Specifications
Total System Specifications
When you use specifications to compare equipment, be sure to establish an error budget that
accounts for all possible measurement uncertainty. Sensor calibration is an integral part of
validation, and total system accuracy should include potential error from the recorder, as well as the
temperature reference and traceable standard. Since all component errors are additive to the total
system, every potential error is significant. A summary of the error budget for a validation system
after sensor calibration with type T thermocouples, used at steam and dry heat, is listed below.
These specifications are guaranteed under worst case conditions. Under typical operating conditions,
you can expect significantly better accuracy.
Kaye Validator (resolution and short term stability)
0.017°C
k=1
IRTD Temperature Standard
0.01°C
k=1
IRTD Temperature Standard
0.051°C
k=1
Total System Uncertainty
0.078°C
k=1
Kaye Validator Specifications
Analog Input
Up to 48
Voltage Input Accuracy
30 days: ±(0.003% of r 2 4
microvolts)
1 year: ±(0.006% of r 2 4
microvolts)
Sensitivity
0.5 microvolts/count on most sensitive range
Voltage Temp. Coef.
±(0.1 micr 0.001% reading)/°C
Compensator Temp. Coef.
±0.01°C per °C
Input Terminal Temperature
Non-uniformity
0.1°C
Input Ranges
-6 to 30mV, -60 to 300mV, -2 to 10V
Environmental
Temperature: 0 to 50°C (32 to 122°F)
Relative humidity: 95% non-condensing
Power
90 to 250 VAC, 50/60 Hz
Fuse Rating
4A Slow Blow
Size
190H X 411W X 381 mm D (457 mm with SIM)
7.5 in H x 16.2 in W x 15 in D (18 in with SIM)
Weight
12 kg (26 lbs)
Battery
Lithium ion 60 minutes for battery backup
150
Kaye Validator User’s Manual
W
LAN
Standards: IEEE 802.11b/g/n
Frequency Range: 2.4 GHz
Data Transfer Rate: IEEE 802.11b: 11, 5.5, 2, 1 Mbps; IEEE 802.11g:
54, 48, 36, 24, 18, 12, 9, 6 Mbps; IEEE 802.11n (1x1 SISO): max.72
Mbps (20 MHz channel); max. 150 Mbps (40 MHz channel)
Summary of Contents for Validator AVS
Page 1: ...Validation Kaye Validator AVS User s Manual Kaye Validator User s Manual...
Page 2: ...Kaye Validator AVS Thermal Process Validation System User s Manual M5100 EN Rev A May 2015...
Page 3: ...No content intended for this page Kaye Validator User s Manual...
Page 133: ...Chapter 10 Generating Reports 126 Kaye Validator User s Manual...
Page 138: ...Chapter 10 Generating Reports Kaye Validator User s Manual 131...
Page 147: ...Chapter 10 Generating Reports 140 Kaye Validator User s Manual...
Page 167: ...No content intended for this page...