Instructions for use ESTETICA E30
7 Safety checks - testing instructions | 7.2 Instructions for safety checks
110 / 134
Using the measuring cable
①
the unit is disconnected from the mains supply
and connection of the treatment centre to the safety tester is enabled. Hence,
the customer-provided mains supply L & N on the power input board need not
be disconnected. The adapter cable
②
is included in the delivery of the KaVo
measuring cable and is required for older treatment centres that are not
equipped with an X2 connector.
Connecting the safety tester to KaVo measuring cables
on the treatment centre
▶
Disconnect plug X2 from the unit control and plug it into the matching con-
nector X2 of the KaVo measuring cable (
Mat. no. 0.411.8811
).
▶
Connect the second plug X2 of the KaVo measuring cable to the unit control
(X2).
▶
Insert the protective contact plug of the KaVo measuring cable into the
sight window.