Instructions for use ESTETICA E30
3 Product description | 3.10 Technical data
34 / 134
3.10 Technical data
Drilling template and setup plan
Layout plan (
Mat. no. 1.009.2781
)
Electrical system
Electrical lead
3 x 1.5 mm
2
(customer-provided
fuse protection 10 A) 3 x 2.5 mm
2
(customer-provided fuse protection
16 A)
Free end above the floor
1 000 mm
Input voltages
100/110/115/120/127/220/230/240
V AC
Frequency
50/60 Hz
Factory-set mains input voltage
See nameplate
Power consumption
100 to 900 VA
Customer-provided fuse protection
Auto-mat C16 or screw-plug fuse 10
A
Protective conductor above floor
See DIN VDE 0100-710, 1000 mm
Heat emission
360 to 3240 kJ/h
Heat emission
Ø 900 kJ/h
Degree of soiling
2
Installation category
II
Mark of approval
CE / DVGW / VDE
Foot control
IPX1: Protection against water drips
Triple function handpiece One
Flush the water and air passages for 20 to 30 seconds before working at the
beginning of the day.
Water pressure
1.5 ± 0.3 bar; flow pressure; probe
tip (1.009.6634)
Water flow
90 ± 15 ml/min
Air pressure
1.5 ± 0.2 bar; flow pressure; probe
tip (1.009.6634)
Air flow
15 ± 3 Nl/min
Triple function handpiece
Flush the water and air passages for 20 to 30 seconds before working at the
beginning of the day.