Instructions for use ESTETICA E30
10 Information about electromagnetic compatibility in accordance with EN60601-1-2 | 10.1 Electromagnetic Trans-
missions
128 / 134
10 Information about electromagnetic compatibility in
accordance with EN60601-1-2
10.1 Electromagnetic Transmissions
The treatment unitESTETICA E30 is designed for operation in an environment
as specified below. The customer or user of the ESTETICA E30 should make
sure that the device is used in an environment of the specified type.
Measurements of emitted in-
terference
Conformance
Electromagnetic environment
-
Guidelines
HF emissions according to CISPR
11
Group 1
The ESTETICA E30 device uses
HF energy for its internal func-
tions exclusively. Therefore, the
HF emission of the device is very
low and interference with adja-
cent electronic devices is unlikely.
HF emissions according to CISPR
11
Class B
The ESTETICA E30 device is suit-
able for use in all facilities includ-
ing residential ones, and facilities
that are directly connected to a
public power supply that also
supplies residential buildings.
Emission of harmonics according
to EN 61000-3-2
Class A
The ESTETICA E30 device is suit-
able for use in all facilities includ-
ing residential ones, and facilities
that are directly connected to a
public power supply that also
supplies residential buildings.
Emission of voltage fluctuations/
flicker according to EN 61000-3-3
Conforms
The ESTETICA E30 device is suit-
able for use in all facilities includ-
ing residential ones, and facilities
that are directly connected to a
public power supply that also
supplies residential buildings.
10.2 Resistance to electromagnetic interference
The treatment unitESTETICA E30 is designed for operation in an environment
as specified below. The customer or user of the ESTETICA E30 should make
sure that the device is used in an environment of the specified type.
Interference im-
munity tests
EN 60601 test level
Compliance level
Electromagnetic en-
vironment -
Guidelines
Electrostatic discharge
(ESD) according to EN
61000-4-2
± 6 kV contact discharge
± 8 kV atmospheric dis-
charge
± 2/4/6 kV contact dis-
charge
± 2/4/8 kV atmospheric
discharge
Floors should be made of
wood or concrete or be
fitted with ceramic tiles.
If the floor is fitted with
synthetic material, the
relative humidity must
be at least 30%.