3 Product description | 3.10 Technical data
38 / 134
Cannula connecting pieces
according to DIN EN ISO 7494 - 2
Diameter of cannula of saliva ejector
and surgical suction
7 mm
Diameter of cannula of spray mist
suction
15 mm
Suction air quantity at spray mist
cannula
minimum 250Nl/min (suction system
with high flow rate), recommended
300Nl/min
Diameter of the suction connection
40 mm
Suction connection above floor
20 mm
Operating environment
WARNING
Inappropriate operating conditions.
Impairment of the electrical safety of the device.
▶
It is essential to comply with the operating conditions specified in the
"Technical Specifications" chapter.
Floor quality
The quality of the flooring must meet
the load bearing ability for buildings
DIN 1055 page 3 and have a pres-
sure resistance in accordance with
DIN 18560 T 1.
Ambient temperature
+10 to +40
o
C
Relative humidity
30 to 75%
Air pressure
700 hPa - 1,060 hPa
Max. elevation for operation
up to 3000 m
Maximum loads
Max. patient weight load
180 kg
Tray holder of the dentist element -
loadable up to
2 kg
Assistant element standard tray -
loadable up to
1 kg
Dentist element - loadable up to
2 kg
Transportation and storage conditions
Ambient temperature
-20 to +55
o
C
Relative Humidity
5% to 95% non-condensing
Air pressure
700 hPa to 1,060 hPa