Instructions for use ESTETICA E30
3 Product description | 3.10 Technical data
36 / 134
WARNING
Risk of infection if the "Water inlet block" is used without additional
safeguards.
Contamination of the treatment water and/or drinking water supply with
germs.
▶
Regarding the use of the "Water inlet block" assembly kit, please note that
no disinfection facility is installed in the unit, and take appropriate safe-
guards.
▶
If the Water bottle assembly kit is used with the enclosed dosing attach-
ment (
Mat. no. 10020287
), add the proper amount of KaVo OXYGENAL 6
(
Mat. no. 04893451
) with each filling. For the correct amount, please
refer to the instructions of the dosing attachment for water germ reduction.
Water germ reduction is integrated in combination with the water bottle with
manual Oxygenal 6 dosing through the dosing attachment and the intensive
germ reduction bottle.
According to DIN EN 1717, each unit that is not listed by DVGW must be
provided with an upstream type AA, AB or AD safety device. (The DVGW water
bottle kit is certified; see the following list.)
When establishing a water connection, make sure that there are no sections of
brackish water with standing water (also in the house plumbing).
For further information, please refer to www.dvgw.de
Free drainage according to DIN EN
1717 - DVGW certified
Water bottle DVGW, reg. no.:
AS-0630BT0111
Water quality
Tap water, cold water connection
Water hardness
1.5 to 2.14 mmol/l
≙
8.4 to 12 °dH
pH
7.2 to 7.8
Customer-site water filtering
80 microns
Water connection
Shut-off valve with brass cone com-
pression screw connection 3/8" to Ø
10 mm provided
Above-floor water connection
min. 50 mm, max. 105 mm with
valve opened
Water inlet pressure
2.0 to 6.0 bar
Water inflow
4 l/min
Diameter of the drain connection
40 mm
Above-floor drain connection
20 mm
Outflow quantity
max. 4 l/min
Slope of water drain pipe
downstream from device: at least 10
mm per metre
Summary of Contents for ESTETICA E30
Page 1: ...Instructions for use ESTETICA E30...
Page 132: ......
Page 133: ......
Page 134: ...1 009 5295 Fk 20180116 10 en...