Instructions for use ESTETICA E30
7 Safety checks - testing instructions | 7.2 Instructions for safety checks
106 / 134
▶
Take the cover off proceeding in upward direction.
7.2.2 Visual inspection (inspection by examination)
Check the following items in advance:
▪ Has the equipment of the ME device or the ME system been changed since
the last inspection?
▪ Was the change documented and approved (test protocol, safety check
STK)?
▪ Are there any indications of insufficient safety?
Check the ratings of fuses that are accessible from
outside
▶
Verify whether the main fuse on the main switch
②
of the unit complies
with the specified nominal data
①
.