Instructions for use ESTETICA E30
2 Safety | 2.2 Purpose – Proper use
12 / 134
Note
The product must be cleaned and serviced according to instructions if it is not
to be used for an extended period of time.
Note
The MULTIflex couplings, the current K/KL motors, and the ultrasonic scaler
hoses of KaVo are equipped as standard with a protective device to prevent
treatment water from being drawn back into the treatment centre via the
handpieces. If products from other manufacturers are used at the standard-
ised interfaces, it must be ensured that they are equipped with an appropri-
ate protective device! If this is not the case, they may not be used!
Information about electromagnetic compatibility
Note
Based on IEC 60601-1-2 (DIN EN 60601-1-2) concerning the electromag-
netic compatibility of electrical medical devices, we must draw your attention
to the following points:
• Medical electrical devices are subject to special precautions concerning the
electromagnetic compatibility and must be installed and operated in accord-
ance with the KaVo assembly instructions.
• High-frequency communications devices may interfere with electrical med-
ical devices.
See also:
2
10 Information about electromagnetic compatibility in accordance with
EN60601-1-2, Page 128
Note
KaVo cannot guarantee the compliance of accessories, cables, and other
components not supplied by KaVo with the EMC requirements of IEC
60601-1-2 (DIN EN 60601-1-2).
Disposal
Note
Any waste which is generated must be recycled or disposed of in strict com-
pliance with all applicable national regulations in a manner which is safe both
for people and the environment.
If you have any questions regarding proper disposal of the KaVo product,
please contact the KaVo branch.
Disposal of electronic and electrical devices
Note
According to EC directive 2012/19 concerning waste electrical and electronic
equipment, this product is subject to the cited directive and must be disposed
of accordingly within Europe.
For more information, please visit www.kavo.com or contact your specialised
dental supplier.
For final disposal: