7 Safety checks - testing instructions | 7.2 Instructions for safety checks
ACP
Protection class 1
WARNING
Electrical power.
Death or injury from electric shock.
▶
Conduct test for leakage current in devices of Protection Class 1 only after
the protective earth test has been passed.
WARNING
Electrical power.
Death or injury from electric shock.
▶
Prior to connecting the treatment centre to the safety tester, disconnect the
treatment centre from the mains supply network.
Measure equivalent patient leakage current
Limit
< 5 mA (maximum value!)
ACP
Protection class 1
WARNING
Electrical power.
Death or injury from electric shock.
▶
Conduct test for leakage current in devices of Protection Class 1 only after
the protective earth test has been passed.
WARNING
Electrical power.
Death or injury from electric shock.
▶
Prior to connecting the treatment centre to the safety tester, disconnect the
treatment centre from the mains supply network.
Note
In the testing of ME devices with several application parts, the parts must be
connected in succession. The measured results must be evaluated using the
threshold values. Application parts, which are not included in the measure-
ment, remain open.
117 / 134