Instructions for use ESTETICA E30
4 Operation | 4.3 Adjusting the dental chair
Move the chair when the instrument is removed
Note
If an instrument is removed, the chair functions of the foot control are
blocked. The blocking can be removed by briefly pressing the stirrup switch.
The functions are then available.
▶
Press the stirrup switch and then the "Preselected spray" or "Blown air" foot
switch.
ð
The chair moves into the selected automatic position.
Saving an automatic position with the foot control
①
Spray preselection/AP footswitch
②
Footswitch blowing air/AP (optional)
③
Foot pedal
The chair positions can be saved on two footswitches; the standard setting is as
follows:
▪ "Spray default" footswitch: "LP" automatic position (last position)
▪ "Blown air" footswitch: "SP" automatic position (rinsing position)
▶
Hold down the foot pedal and foot-operated button "SP", and simultan-
eously press any button for an automatic position ("AP 0", "AP 1", "AP 2" or
"SP") on the dentist or assistant unit until you hear a beep.
ð
The automatic position is saved to the foot-operated button.
or
▶
Hold down the foot pedal and foot-operated button "LP", and simultan-
eously press any button for an automatic position ("AP 0", "AP 1", "AP 2" or
"SP") on the dentist or assistant unit until you hear a beep.
ð
The automatic position is saved to the foot-operated button.