Instructions for use ESTETICA E30
7 Safety checks - testing instructions | 7.2 Instructions for safety checks
Visual inspection and appraisal of the medical device
and accessories
The following list is for exemplary purposes and makes no claim of being com-
plete.
Check the following items:
▪ Stability of the device
▪ No damage to the cladding or casing (cracks, breakage)
▪ Functioning of the carrier systems on dentist and assistant side, treatment
lamp, and display (brakes, height adjustment, etc.)
▪ Condition of the handpiece and suction hoses
▪ Condition of all installed application parts
▪ Condition of the control panels
▪ Condition of the threads for the fitting of tips to the ultrasound scaler hand-
piece
▪ Condition of the operating light
▪ Absence of leaks on the body of the device
▪ Condition of the power connection provided by the treatment centre
▪ Condition of air and water connections
▪ Any damage on the sight window and the casing of the camera ERGOcam
▪ Expiry date of the water bottle inserted in the BS water bottle not exceeded
Check of legibility and completeness of the safety-
related labels
▶
Check if all safety-related markings (plates and labels) are present and
legible.
▶
Check if the rating plate and serial number plates are present and legible.