4 Operation | 4.11 Using instruments
88 / 134
Note
If no patient is selected, the images stay in the "Swap Tray" and are not
saved permanently. As soon as a patient is selected, these temporary data in
the "Swap Drying" are deleted. When an active patient is logged off (or a
new patient is logged on) in CONEXIO, a query is shown asking whether the
images shall be deleted or saved. Data deleted at this point cannot be re-
stored subsequently.
4.11 Using instruments
Note
Consult the pertinent separate instructions for information regarding the in-
stallation, use and servicing of the individual handpieces (such as the turbine,
PIEZOsoft, Satelec Mini LED, etc.).
4.11.1 Setting spray air and spray water
The adjusting screws for spray air and spray water are situated on the under-
side of the table and correspond to the handpiece that is actually connected.
For KaVo Midwest handpieces, spray air and spray water are pre-set suitable at
the factory such that no re-adjustment is required.
3
2
1
▶
Use the short adjusting screw
②
on the control valve to adjust the spray
air.
ð
Counterclockwise rotation (seen from below) increases the flow rate.
▶
Use the long adjusting screw
①
on the control valve to adjust the spray wa-
ter.
ð
Counterclockwise rotation (seen from below) increases the flow rate.