Instructions for use ESTETICA E30
3 Product description | 3.10 Technical data
Water pressure
1.5 ± 0.3 bar; Flow pressure; 4 x
manometer
Max. static pressure water
2.5 ± 0.3 bar
Water flow
80 ± 10 ml/min
Air pressure
3.3 ± 0.1 bar; Flow pressure; 4 x
manometer
Max. static pressure air
4 + 0.5 bar
Air flow
at least 16 Nl/min
Water supply
CAUTION
The ESTETICA E30 treatment centre, as delivered, is not equipped
with an amalgam separator.
Contamination of wastewater.
▶
Disposing wastewater, it is imperative to comply with the national regula-
tions, in particular with regard to amalgam loads.
Note
KaVo recommends to connect an external amalgam separator to the ESTET-
ICA E30 treatment centre in compliance with the specifications of the manu-
facturer of the device.
Note
If the water is very hard (above 12 °dH), a water softening device must be
fitted in the ion-exchange process.
Insufficient water hardness (below 8.4 °dH) can promote the formation of al-
gae.
Note
The "water inlet block" assembly kit does not include a separation between
the treatment water and water supplied by the local mains. The operator
must observe and comply with relevant national directives concerning the
prevention of backflow. Upon failure to comply with these rules, the manu-
facturer accepts no liability for the quality of the treatment water and the mi-
crobial re-contamination of the public drinking water network.
WARNING
There is a risk of infection if you fail to comply with national regula-
tions.
Contamination of the treatment water and/or drinking water supply with
germs.
▶
Note and comply with national regulations concerning the quality of water
for human use (drinking water) - if applicable.
▶
Note and comply with national regulations concerning the prevention of
backflow (flow from the treatment centre back to the public water supply) -
if applicable.