Instructions for use ESTETICA E30
4 Operation | 4.11 Using instruments
Borden handpieces
Note
In the case of Borden handpieces, the drive air needs to be adjusted for each
handpiece by a Service Technician using adjusting screw
③
. The adjustment
can be made using the measuring set
Mat. no. 0.4179807
(
Mat. no.
10039374
) or the test manometer for Borden handpieces
Mat. no.
04118531
.
4.11.2 Using suction hoses
▶
Remove the spray mist suction device and/or saliva ejector from the holder.
ð
The spray mist suction device and/or saliva ejector is turned on automatic-
ally and turned off when it is replaced in the holder.
The suction flow of the saliva ejector or spray mist suction device can be re-
duced or blocked with the slide valves integrated in the handpieces.
5
1
2
3
4
5
6
6
▶
Move the slide valve completely upward.
ð
The slide valve is open: maximum suction.
▶
Move the slide valve down all the way.
ð
The slide valve is closed: no suction.
Note
Connectors for the spray mist suction and the saliva ejector without slider as
well as reducing pieces for the spray mistsuction are available as accessories.
▪ Cannula holder, short, for spray mist ejector
②
(
Mat. no. 0.764.5783
)
▪ Cannula holder, long, for spray mist ejector
②
(
Mat. no. 0.764.5853
)
▪ Cannula holder, short, for saliva ejector
①
(
Mat. no. 0.764.5863
)
89 / 134