JIF-2002 WEIGHING INDICATOR&CONTROLLER
45
5. When the display reaches "
Final - FREE
", the FREE Output (Pin 10) signal
will come
ON
. Closed the Dribble-Flow Gate by using the FREE Output ON
signal.
6. After Free Fall has stopped - check if the HI and LO (Pin 9, 8) signals are
OFF. If both outputs are OFF then the batch is completed correctly.
7. An Automatic Free Fall Compensation Command (Min. 200ms pulse to Pin
21) may be given at this time. If you change the Free Fall Set Point value
either from the Front Panel or the RS-232C, RS-422/485
-
the learned
Free Fall value will be cleared.
8. Use the FREE (Pin 10) signal to delay a time period as the control signal is
processing empty the Weighing Hopper.
9. When the GROSS weight is below the ZERO band, the ZERO Band Output
will come ON -signifying the Weighing Hopper is empty. Closed the
Weighing Hopper Discharge Gate by using the ZERO Band (Pin 13) Output
ON signal.
10. You are now ready for your next batching event.