MP73P+ : Battery charger/analyser
Operators Manual
CAP TEST CELL MONITOR
Value Required
Units
Description
Selection
-
Select either : -
TIME & TARGET = 100% OR LESS
TARGET ONLY = ACTUAL CAPACITY
Number of Cells
1 to 20
The number of cells in the battery.
Target Volts
Volts.
This is the minimum cell voltage that the discharge
must go to. Once one cell meets this voltage the
capacity test will stop, if before the discharge time.
Once it stops it will report the battery capacity.
Discharge Amps
Amps
The discharge current to maintain.
Discharge Time
Hour and
Minutes.
This is the discharge time (usually 1 hour). If TIME &
TARGET has been selected when this time is reached
the charge will stop and report the battery capacity of
100%, otherwise the discharge will continue to report
actual capacity.
Cell Print Interval Minutes.
During the whole charge the cell voltages can be
printed at periodic intervals in minutes. If not required
set to zero. Note during the charge the cell voltages
can be manually printed at any time by pressing the '.'
key.
CAPTEST TO FAILED CELLS
Value Required
Units
Description
Selection
-
Select either : -
TIME & TARGET = 100% OR LESS
TARGET ONLY = ACTUAL CAPACITY
Number of Cells
1 to 22
The number of cells in the battery.
Target Volts
Volts.
The battery voltage at which point the discharge is to
stop if the time up has not elapsed. At which point it
will report the battery capacity as a function of time.
Discharge Amps
Amps
The discharge current to maintain.
Discharge Time
Hour and
Minutes.
The maximum time for the capacity test usually 1
hour. When this time is elapsed the battery capacity
will be reported as 100%.
Cell Fail Volts
Volts.
During the discharge the cell voltages are monitored
and if any cell voltage falls below this figure the cell
Doc: DWG1042-12-R18 MB73P+ Operators manual.odt
Page 45 of 64