Operators Manual
•
A confirmation screen will be displayed and the unit will go back to the Main
Menu.
Note the unit has a build-in calibration checking and if any of the entered data does not
correlate to valid calibration values stored internally it will sound the alarm and display
a warning message and the calibration will be discarded. This protects the unit from
incorrectly entered values.
The calibration procedure will have to be re-executed and if the error is repeated then
the unit may have a fault and will need to be returned for service or repair.
10.5
Reviewing Calibration Date
The calibration parameters can be displayed or printed out on the built in printer. To
perform this operation the following procedure must be observed.
Select
SETTINGS
->
or
DISPLAY
->
SYSTEM
The unit will then print or display the calibration data.