MP73P+ : Battery charger/analyser
Operators Manual
8.3
Entering Discharge Data
From the
MAIN
screen by selecting
CHARGE
->
MANUAL
From the
MAIN
screen by selecting
CHARGE
->
LIBRARY
->
->
MODIFY
From the
MAIN
screen by selecting
PROCESS
->
PROCESS #
->
VIEW/EDIT
->
MODIFY
->
CHARGE BATTERY
.
The screen will request
CHARGE TYPE
.
Select
>> NEXT >>
or
<< PREV <<
to change to either : -
STANDARD CAPACITY TEST
CAP TEST CELL MONITOR
CAPTEST TO FAILED CELLS
CAPTEST FIXED RESISTOR
AUTOMATIC CELL BALANCE
FULL DISCHARGE
The following tables explains what the values entered mean.
STANDARD CAPACITY TEST
Value Required
Units
Description
Selection
-
Select either : -
TIME & TARGET = 100% OR LESS
TARGET ONLY = ACTUAL CAPACITY
Nominal Volts
Volts.
The battery nominal voltage is used to determine what
power rating is required for discharge and also the
help validate the data entered.
Target Volts
Volts.
The target volts is the discharged voltage that the
charge should stop the discharge and report the
battery capacity.
Discharge Amps
Amps.
This is the constant current discharge amps that the
battery has to be discharged at.
Discharge Time
Hours and
Minutes.
This is the discharge time (usually 1 hour). If TIME &
TARGET has been selected when this time is reached
the charge will stop and report the battery capacity of
100%, otherwise the discharge will continue to report
actual capacity.