• There are no electrical connections from
the tonometer to the patient.
• The device has B-type electric shock
protection.
• Operation environment:
• Temperature: +10 °C to +35 °C
• Relative humidity: 30 % to 90 %
• Atmospheric pressure: 800 hPa-1,060 hPa
• Storage environment:
• Temperature: -10 °C to +55 °C
• Relative humidity: 10 % to 95 %
• Atmospheric pressure: 700 hPa-1,060 hPa
• Transport environment:
• Temperature: -40 °C to +70 °C
• Relative humidity: 10 % to 95 %
• Atmospheric pressure: 500 hPa-1,060 hPa
• Mode of operation: continuous.
PERFORMANCE DATA
The performance data is obtained from
a clinical study, performed according to
American National Standard ANSI Z80.10-
2003 and International Standard ISO 8612.2
for tonometers. The study was performed ain
the Department of Ophthalmology, Helsinki
University Central Hospital. In the study,
158 patients were measured. The mean paired
difference and standard deviation (Goldmann-
Icare) were -0.4 mmHg and 3.4 mmHg.
A scattergram and Bland-Altman plot of the
results is shown in figure 14.
ACCESSORIES
SKU
PRODUCT DESCRIPTION
WEIGHT DIMENSIONS
104
BOX OF 100 PROBES
89,10 g
5,3 x 10,3 x 3,6 cm
102
BOX OF 100 PROBES (ONLY USA)
89,10 g
5,3 x 10,3 x 3,6 cm
7210
PROBE BASE COLLAR
1g
11 mm x 10 mm
540
PROBE BASE
4g
7 mm x 38 mm
550
TABLE STAND
52g
73 mm x 52 mm x 38 mm
560
WRIST STRAP
3g
270 mm x 10 mm x 10 mm
500
ALUMINIUM CASE
700g
240 mm x 280 mm x 72 mm
7171
BATTERY COVER & SCREW
4g
42 mm x 17 mm x 13 mm
7000
THREAD PLATE
1g
4 mm x 10 mm x 3 mm
543
PROBE BASE CLEANING CONTAINER 3g
5,6 cm x 2 cm
561
562
563
564
SILICON GRIP - WHITE
SILICON GRIP - PINK
SILICON GRIP - GREEN
SILICON GRIP - BLUE
26g
135 mm x 46 mm x 23 mm
MAINTENANCE
Follow local regulations and recycling
instructions regarding the disposal or recycling
of the Icare tonometerand accessories.
WARNING
The tonometer should only be opened by
qualified service personnel. It contains no
userserviceable parts, apart from the batteries
and a probe base. The Icare tonometerrequires
no routine servicing or calibration other than
changing the batteries at least every 12 months
or changing or cleaning the probe base. If
servicing is necessary, contact qualified service
personnel or your local Icare representative.
REPLACING THE PROBE BASE
Replace the probe base every twelve months.
Clean or replace the probe base if the error
messages E01 or E03 are displayed.
Instructions for replacing the probe base:
•
Turn off the tonometer.
• Unscrew the probe base collar and put it in a
safe place.
• Remove the probe base by tilting the
tonometer downwards and use your fingers
to pull the probe base out of the tonometer.
• Insert a new probe base into the tonometer.
• Screw the collar in, to lock the probe base.