8
Icare tonometer TA01i manual
GUIDANCE AND MANUFACTURER’S DECLARATION-ELECTROMAGNETIC EMISSIONS
TA01i is intended for use in the electromagnetic environment specified below. The user of the
TA01i should assure that it is used in such an environment.
RF emissions CISPR 11
Group 1
Icare TA01i is battery operated and use RF
energy only for its internal function. Therefore,
its RF emissions are low and are not likely to
cause any interference in nearby equipment.
RF emissions CISPR 11
Class B
Icare TA01i is suitable for use in all
establishments, including domestic
establishments and those directly connected to
public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
NOT APPLICABLE
Voltage fluctuations
flickering emissions
IEC 61000-3-3
NOT APPLICABLE
GUIDANCE AND MANUFACTURER’S DECLARATION-ELECTROMAGNETIC IMMUNITY
Icare TA01i is intended for use in the electromagnetic enviroment specified below. The customers
or users of Icare TA01i should assurre that it is used in such enviroment
Immunity test
IEC 60601
Test level
Compliance
level
Electromagnetic environment-
Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%
Electrical fast
Transients/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
NOT
APPLICABLE
Surge
IEC 61000-4-5
±1 kV for line(s) to
line(s)
±2 kV for line(s) to
earth
NOT
APPLICABLE
Voltage dips, short
interruption and
voltage variations on
power supply lines
IEC 61000-4-11
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 s
NOT
APPLICABLE
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital environment.