ibidi Pump System
• Immediately replace damaged cords, plugs, or cables to avoid risk of personal injury or damage
to the instrument.
• Only ibidi technical staff and technical staff instructed by ibidi are permitted to open and ser-
vice the ibidi Pump System.
• The external power supply should not be brought into contact with moisture. If the housing is
damaged, the external power supply should not be used.
• Avoid strong magnetic fields and sources of high frequency. The ibidi Pump System might not
function properly when located near a strong magnetic field or high frequency source.
• Avoid vibrations from vacuum pumps, centrifuges, electric motors, processing equipment, and
machine tools.
• Avoid dust and corrosive gas. Do not install the ibidi Pump System where it could be exposed
to high levels of dust or to outside air or ventilation outlets.
• Install the ibidi Pump System in a location that enables easy access for maintenance.
• Do not place heavy objects on the instrument.
• The weight of the ibidi Pump is approx. 3 kg. Moving the pump during operation will pose a
risk of personal injury or damage to the instrument.
• The ibidi Pump can build pressure up to 100 mbar. Do not unplug the fluidic connections dur-
ing pump operation. Pressurized liquid could emerge from the tubes and damage surrounding
equipment. Excess moisture can cause the external power supply or nearby electrical equip-
ment to short circuit.
• Do not suction any liquid into the ibidi Pump.
Instruction Manual
11
Version 2.6 (May 12, 2022)