3-32
i-CAT FLX Installation Manual
032-0330-EN Rev K
Installation Check for a DEXIS i-CAT FLX Installation
1.
Start DEXIS.
2.
In DEXIS, enter data for a test patient.
a.
Select
New Patient
.
b.
Enter data for the patient and exam. Example: Enter
Test
for Last Name and
Patient
for
First Name.
c.
Select
OK
.
d.
Click
(Extra-oral) and click
I Accept
on disclaimer pop-up message.
e.
Click .
Click
Later
on registration pop-up message.
f.
A pop-up message displays that patient is scheduled for acquisition. Click
OK
to dismiss.
3.
In SmartScan STUDIO, capture a scan for test patient.
a.
Set up a phantom in the scanner.
b. At scanner controller, access
Acquire
option.
c.
Select
Test Patient
from Scheduled Exams list and press
.
d.
Select a protocol and press
.
e.
Take a scan.
f.
Review image and press
to complete.
4.
In DEXIS, locate Test Patient exam in Extra-oral display.
5.
Double-click the i-CAT box to start TxSTUDIO and load exam.
6.
View scan in TxSTUDIO and ensure correct path to the TxSTUDIO Cases
folder displays in File Manager.