3-24
i-CAT FLX Installation Manual
032-0330-EN Rev K
e.
Check the status indicators at the top of the main window. All
indicators should display
OK
. If not, go to
SmartScan STUDIO
Manager Status Indicators (page
3-6)
and troubleshoot.
f.
Click
and close SmartScan STUDIO Manager.
11. For DEXIS sites, check settings on DEXIS i-CAT FLX Configuration:
a.
Open Control Panel and select
DEXIS
i-CAT FLX
.
b.
Check Integration Services Connection
options.
Automatic By Name
(default)
connects to SmartScan STUDIO
Integration Services server location by
computer name. This prevents
connection problems caused by
changing IP addresses. Some network
setups may not be able to resolve the
computer name, and will have to use
an IP address (for example, computers
on different subnets).
When required, select
Manual By Address
option to specify the server running
SmartScan STUDIO Integration Services by IP address and Port (must match address/
port in SmartScan STUDIO Integration Services Configuration window). A fixed IP
address should be used. See
Fixed IP Addresses (page
3-8)
for additional information.
c.
Check that Scanner and Image Database Folder locations match what is displayed in
SmartScan STUDIO Integration Services Configuration window. Click
OK
.
Install TxSTUDIO Software
NOTE:
If TxSTUDIO software is already loaded and configured on all workstations requiring it, skip this
section.
TxSTUDIO should be loaded on the workstation where the site will be viewing QA Tests. Additionally,
it should be loaded on workstations that will be using it to view patient images, as purchased by the site.
For sites that want to share workups among workstations, a shared folder must be created on the long-
term storage server and TxSTUDIO configured so that all workups will be saved to this location. This
will ensure that all users can access workups from any workstation that has TxSTUDIO loaded.
1. At the long-term storage server, create a folder named
TxSTUDIO Cases
that can be shared
by all the site workstations that will use TxSTUDIO. Ensure that all workstations running
TxSTUDIO have read/write permissions to this location. Coordinate with site IT personnel if
necessary.
2.
To download TxSTUDIO from the website:
a.
Ensure there is an internet connection at the workstation. Insert media in drive where
downloaded files will be copied.
NOTE:
You can also download TxSTUDIO to a shared network location.
b.
Go to
www.anatomage.com
, then click on
Dr. Login
. Contact Technical Support for
username and password, and also for the authorization code(s) for each copy of
TxSTUDIO to be loaded.
c.
Download the latest TxSTUDIO installer to the media and remove media.
Summary of Contents for FLX
Page 6: ... vi i CAT FLX Installation Manual 032 0330 EN Rev K ...
Page 26: ...1 18 i CAT FLX Installation Manual 032 0330 EN Rev K ...
Page 84: ...4 14 i CAT FLX Installation Manual 032 0330 EN Rev K ...
Page 86: ...5 2 i CAT FLX Installation Manual 032 0330 EN Rev K ...
Page 104: ...C 10 i CAT FLX Installation Manual 032 0330 EN Rev K ...
Page 105: ......