3-17
Load Clinical Software and Configure System
032-0330-EN Rev K
Install and Configure Clinical Software
Install SmartScan STUDIO Server Components
SmartScan STUDIO Integration Services should be loaded on the server that will be used as the long-
term data storage device. Load Data Utilities, Legacy Practice Management Interface, and PACS
Storage Service as needed by the site.
NOTE:
•
During installation, acknowledge any Microsoft User Account Control notifications that may be
displayed to continue installation.
CAUTION
CAUTION
Be sure to load SmartScan STUDIO Integration Services on only one workstation at a site, which
should be the server that will be used as the long-term data storage device. Workflow at client
workstations will be impacted if an invalid Integration Services server is configured.
1. At server, insert media in drive.
2.
On AutoPlay pop-up, select
Run SssInstallerExecutive.exe
. If AutoPlay pop-up does not
display, navigate to the media drive, right-click and select
Open AutoPlay
.
3.
On the SmartScan STUDIO Installer Executive:
a.
Select programs to be installed.
•
Integration Services (required)
•
PACS Storage Service - (for sites using a PACS for image storage)
•
Data Utility (optional)
•
Patient Data Utility (optional)
•
Legacy Practice Management Interface - (for sites using a supported PM system)
•
DEXIS
i-CAT FLX Server Plugin - (for sites using DEXIS)
Summary of Contents for FLX
Page 6: ... vi i CAT FLX Installation Manual 032 0330 EN Rev K ...
Page 26: ...1 18 i CAT FLX Installation Manual 032 0330 EN Rev K ...
Page 84: ...4 14 i CAT FLX Installation Manual 032 0330 EN Rev K ...
Page 86: ...5 2 i CAT FLX Installation Manual 032 0330 EN Rev K ...
Page 104: ...C 10 i CAT FLX Installation Manual 032 0330 EN Rev K ...
Page 105: ......