2-5
Calibration and Laser Adjustments
032-0330-EN Rev K
Rotate Crosshair Laser Lines:
7.
Check crosshair lines with panel cover notches
to see if a rotation adjustment is required.
7
8.
9
8
To rotate laser crosshairs slightly loosen the two
laser mounting screws (shown).
9.
To help rotate the laser, insert an allen wrench
(3/32) into the top hole of the laser.
10. Tighten both screws when properly aligned and
remove 3/32 allen wrench.
11. Replace beam limiter cover.
Patient Chair Alignment
Install Chair Calibration Fixture
1.
Insert chair calibration fixture into positioning block.
2.
Using the
Alignment Light
, check position of the
horizontal and vertical lasers relative to the notches on the
chair calibration fixture, as shown.
•
If horizontal laser is out of position, move chair
calibration fixture up or down in the positioning block,
and pull back on the fixture if needed to roughly align
with horizontal notch. Tighten knob.
V
ertical
Laser
Horizontal
Laser