1-10
i-CAT FLX Installation Manual
032-0330-EN Rev K
47.
47
48
49
3/32
Slide inner source cover into place, onto
top
mounting notches
.
48. Mount outer cover onto mounting notches, over panel
cover.
49. Attach covers with hex mounting screws at top and
bottom (quantity 4 outer cover and 4 inner source
cover).
50.
56
57
50
51
52
53
Connect chair control cable to
1520
(Chair) connector
on rear of overhead panel.
51. Connect control box cable to
1518
(Control Box)
connector on rear of overhead panel. Connectors are
keyed to prevent improper insertion.
For wall mounting options, see
Wall Mount Control
Box
chapter.
52. Connect shorting plug to
1525
(door interlock) on rear
of overhead panel.
53. Connect Ethernet cable for scanner controller to rear
of overhead panel.
54.
55
54
Connect chair control cable to patient chair
connector.
55. Connect patient emergency stop control to patient
chair connector. Connectors are keyed to prevent
improper insertion. Rotate red emergency stop button
to verify that it is disengaged.