3-23
Load Clinical Software and Configure System
032-0330-EN Rev K
8.
If the DEXIS i-CAT FLX Client was selected to be installed:
a.
On DEXIS i-CAT FLX Configuration screen, click
Next
.
b.
On Ready to Install screen, click
Install
. Installation will start and a progress bar is
displayed.
c.
When installation completes, click
Finish
on the Install Shield Wizard screen.
9.
On SmartScan STUDIO Installer Executive, check that all selected programs have been
installed, then click
Close
. Remove media.
10. Launch SmartScan STUDIO Manager from desktop:
a.
Select
to access Settings.
b.
Select image viewer to be used on the workstation.
c.
Check that Scanner and Image Database Folder locations match what is displayed in
SmartScan STUDIO Integration Services Configuration window.
d.
Check Integration Services Connection options.
•
Automatic by Name
(Default) - This is the preferred option. Connects to SmartScan
STUDIO Integration Services server location by computer name. This prevents
connection problems caused by changing IP addresses. This field must match the name
in SmartScan STUDIO Integration Services Configuration window. Some network setups
may not be able to resolve the computer name, and will have to use an IP address (for
example, computers on different subnets).
•
Manual by Address
- When required, select this option to specify the server running
SmartScan STUDIO Integration Services by IP address and Port (must match address/
port in SmartScan STUDIO Integration Services Configuration window). A fixed IP
address should be used. See
Fixed IP Addresses (page
3-8)
for additional information.
Summary of Contents for FLX
Page 6: ... vi i CAT FLX Installation Manual 032 0330 EN Rev K ...
Page 26: ...1 18 i CAT FLX Installation Manual 032 0330 EN Rev K ...
Page 84: ...4 14 i CAT FLX Installation Manual 032 0330 EN Rev K ...
Page 86: ...5 2 i CAT FLX Installation Manual 032 0330 EN Rev K ...
Page 104: ...C 10 i CAT FLX Installation Manual 032 0330 EN Rev K ...
Page 105: ......