2-7
Calibration and Laser Adjustments
032-0330-EN Rev K
Centering Patient Chair:
8.
Select
90
o
and press
.
9.
When prompted, press
Scan
button on control box. An audible alarm is sounded and the X-ray
ON light is illuminated during radiation exposure.
10. Review the scout image.
11. Verify that the vertical pin falls within the double lines. If
aligned, go to step 14.
12. If the pin is not within the double lines:
a. Loosen 4 screws in rear of scanner.
b. Move chair left or right by pushing on side of chair.
Alternately, insert screwdriver in rear slot of scanner and
pry tool left or right depending on adjustment needed.
NOTE:
Typically, to move the pin left, move chair
assembly right (when facing front of scanner). To move
the pin right, move chair assembly left (when facing front
of scanner).
13. Repeat steps 8-12 as needed until pin is within the double
lines on scout scan.
Use
as needed to page through scout scans to help
with adjustment.
14. When pin falls within the double lines at both 0
o
and 90
o
,
tighten chair mounting screws.
15. After chair assembly mounting screws are tightened, repeat ChairCal at 0
o
and 90
o
to verify
chair alignment.
16. If the vertical pin falls outside the double lines at either position, repeat the necessary steps to
adjust the chair, and perform a scout, until chair alignment is verified.
Adjust Centerline Laser
WARNING
Do not stare into laser. Severe personal injury (blindness) may result.
NOTE:
Ensure that the patient chair alignment has been completed before beginning this
procedure.
12
Summary of Contents for FLX
Page 6: ... vi i CAT FLX Installation Manual 032 0330 EN Rev K ...
Page 26: ...1 18 i CAT FLX Installation Manual 032 0330 EN Rev K ...
Page 84: ...4 14 i CAT FLX Installation Manual 032 0330 EN Rev K ...
Page 86: ...5 2 i CAT FLX Installation Manual 032 0330 EN Rev K ...
Page 104: ...C 10 i CAT FLX Installation Manual 032 0330 EN Rev K ...
Page 105: ......