2-10
i-CAT FLX Installation Manual
032-0330-EN Rev K
5.
If the Geometric Calibration Processing Failure message is displayed, the detector pivot is out
range. The error message displays the allowed value (+/-0.050 degrees) and the actual value.
To adjust detector pivot in landscape position:
a.
If receptor cover is installed, remove both mounting screws (3/32”) from the bottom and
remove cover from mounting tabs, which are located at the top.
b.
On the receptor assembly, loosen the 5/32” set
screw.
c.
The 1/4” adjustment screw is used to adjust the
detector pivot.
Turning the adjustment screw one full turn is an
adjustment of
0.3
.
Typically, turn screw
counter-clockwise
to
subtract
from the displayed number and
clockwise
to
add
to the displayed number
(facing the screw).
d.
A hex shape consists of six flat sides. Measure
the turns by hex flats which equal
0.05
per flat.
Six flats equal
0.3
.
e.
Tighten the 5/32” set screw after the adjustment
is made.
f.
In GeoCal, press
OK
to rerun in landscape
position. Repeat as needed until landscape
calibration passes.
6.
Press
to initiate portrait calibration. The
panel rotates.
7.
When prompted, press the
Scan
button on the operator control box. An audible alarm is
sounded and the X-ray ON light is illuminated during radiation exposure.
8.
If the portrait calibration passes, the message
Step 2 Complete
is displayed. Go to step 10.
9.
If the Geometric Calibration Processing Failure message is displayed, the detector pivot is out
of range. The error message displays the allowed value (+/-0.050 degrees) and the actual
value. To adjust detector pivot in portrait position:
5b
5c
Summary of Contents for FLX
Page 6: ... vi i CAT FLX Installation Manual 032 0330 EN Rev K ...
Page 26: ...1 18 i CAT FLX Installation Manual 032 0330 EN Rev K ...
Page 84: ...4 14 i CAT FLX Installation Manual 032 0330 EN Rev K ...
Page 86: ...5 2 i CAT FLX Installation Manual 032 0330 EN Rev K ...
Page 104: ...C 10 i CAT FLX Installation Manual 032 0330 EN Rev K ...
Page 105: ......