HYDAC FILTER SYSTEMS GMBH
Carrying out the maintenance / inspection | 7
BeWa CTU1x3x-CTU1x4x 4093420f en-us web
93 / 124
Filling the CTU1x4x/CTU1x5x
To fill the extraction unit/to top up the test liquid, proceed as fol-
lows:
ü
Have the container with the test liquid ready.
5. Detach the plug with the breather filter from the ball valve.
6. Attach the filling hose to the ball valve and turn the ball valve
to the
Vent
position.
7. Start the
Fill system
menu item in the user interface.
ð
The extraction unit is filled. Overfilling is not possible, the
software automatically stops filling after the maximum filling
quantity has been reached.
8. Turn the ball valve in Operation position.
9. Remove the filling hose from the ball valve.
10. Attach the plug with the breather filter to the ball valve.
O
The filling of the extraction unit is complete.