Fig. 37:
Preparing the protective gloves
Insert the protective gloves through the inlet opening on the analy-
sis chamber. Then stretch the led end of the glove over the inlet
opening from the outside. It is not necessary to additionally secure
the protective gloves to the inlet opening.
Using electrostatically conductive protective gloves
When replacing the protective gloves, ensure that there is suffi-
cient electrostatic conductivity as specified in DIN EN 61340-5-1.
R
g
= 2.0 x 10
8
Ω
Make sure that the resistor R
g
(Ω) to EPA ground or to a grounding
point is between 7.5x10
5
≤ R
g
≤ 1x10
12
for protective gloves and
fingerstalls.
Removing the protective gloves
To remove the protective glove, pull it off the rolled end of the inlet