6-16
HP 85106D mm-Wave System
HP 85105A mm-Wave Controller
Specifications
Specifications
Specifications describe the warranted performance of the instrument. The
electrical specifications of the HP 85105A mm-wave controller when used in
an HP 85106D system are in Chapter 3, “Performance Verification” in this
manual.
Supplemental Characteristics
The supplemental characteristics provide information useful in applying the
instrument by giving typical, but non-warranted, performance parameters.
NOTE
The RF level can be adjusted more than 10 dB down from nominal by use of
the Source 1 power level menu on the HP 8510.
Table 6-1
HP 85105A Supplemental Characteristics
Test Ports
RF and LO Connector Type:
Precision 3.5 mm male
Connector Torque:
90 N-cm (8 in-lb)
Nominal Operating Power Level
Operating Level
Port 1
Port 2
RF 8 to 20 GHz
>+ 17 dBm
>+ 17 dBm
20 GHz to 21 GHz
(Option 050, or 054)
>+ 16 dBm
LO 2 to 8 GHz
>+ 20 dBm
>+ 20 dBm