HP 85106D mm-Wave System
6-35
HP 85105A mm-Wave Controller
RF Amplifier Output
1. Run the system software and select {
HIGHBAND
} to make the mm-wave
controller active.
NOTE
Make sure the highband test set is not connected to the module interface
connectors.
2. Choose a power meter that covers the 13 GHz to 22 GHz frequency
range (HP 436A meter and HP 8485A sensor). Calibrate the power
meter and attach a 3.5 mm 10 dB fixed attenuator (HP 8493C) and a
female to female adapter to the sensor.
3. Loosen the end of the coax cable attached to the output of the RF
leveling amplifier coupler and detach the other end of the coax cable
from the pin switch. Rotate the cable and attach the power meter sensor.
Tighten the cable end attached to the RF leveling amplifier. Select
{
SINGLE POINT MODE
} from the [
STIMULUS
] menu, then using the RPG
knob, slowly scan over the frequency band. Power should be
+22 dBm
±
2 dB.