HP 85106D mm-Wave System
3-3
Performance Verification
Installation of HP-BASIC and Performance Verification Software
Installing on an
HP 300 Series
Controller
The performance verification software runs on most HP 200 and HP 300
series (except for HP 9826) controllers and requires at least two megabytes
of memory to run the program.
1. Load HP BASIC (rev 5.0 or later), and make sure the following
language extension and driver files are installed one at a time.
•
ERR
•
CLOCK
•
GRAPH
•
MAT
•
IO
•
HP-IB
Insert the HP 8510C Specification and Verification software disk
(LIF format) into the default drive (or the drive you specify as the MSI
(mass storage is).
•
Load the file by typing:
LOAD “SPECS_8510”
, then press [
RETURN
] or [
EXECUTE
]
2. Type:
RUN
This will start the program.
For Both PC and
HP 300 Controller
After the program is started:
1. The display allows you to set the date and time.
At this point, you may press the controller “Y”' key to continue, or you press
the controller "N" key and set the date and time. When your entries are
complete, press [
RETURN
] or [
ENTER
].
2. The program loads the System Configuration file and System Hardware
Configuration Menu is displayed. Use this menu to select the system
equipment you want to use.
Summary of Contents for 85106D
Page 11: ...x HP 85106D mm Wave System ...
Page 117: ...4 24 HP 85106D mm Wave System Service and Troubleshooting Check Unratioed Power Levels ...
Page 139: ...5 22 HP 85106D mm Wave System HP 85104A Test Set Module Troubleshooting Procedures ...
Page 140: ......
Page 141: ......
Page 142: ......
Page 143: ......
Page 208: ......
Page 238: ...B 2 HP 85106D mm Wave System Upgrade Paths ...
Page 243: ...HP 85106D mm Wave System v work surface installation 1 12 ...
Page 244: ... vi HP 85106D mm Wave System ...