![HP 85106D Manual Download Page 134](http://html.mh-extra.com/html/hp/85106d/85106d_manual_161501134.webp)
HP 85106D mm-Wave System
5-17
HP 85104A Test Set Module
Troubleshooting Procedures
Check the RF Power
Out of A13 Source
Block
Procedure 7
The following equipment is required to perform this procedure:
Before beginning this procedure, turn the HP 85105A mm-wave controller
power OFF. Disconnect the HP 85104A from the HP 85105A. Refer to step
references in Figure 5-11 on page 5-32.
1. Disassemble the test-set module using the following four steps:
a.
Invert the test set module so that the bottom aide is facing up.
Remove the stand by loosening the two captive screws until the
module and stand are disengaged.
b.
Remove all extensions from the HP 85104 test set modules.
c.
Remove the two torx-head screws from the bottom housing. Pull the
bottom housing off.
2. Disconnect the LO IN and LO OUT cables from the power divider
(Step 13).
3. Remove the two screws from the power divider to remove the power
divider (Step 17).
4. Disconnect the main line isolator (A7, Step 18) from the source and
coupler interfaces.
5. Connect the power sensor to A13 module assembly output as shown in
Equipment
Required
Q-band
U-band
V-band
W-band
Power Meter
HP 436A
Anritsu ML83A
Anritsu ML83A
HP 436A
HP 436B
HP 438A
Power Sensor
HP Q8486A
Anritsu MP715A-004
Anritsu MP716A
HP W8486A
Summary of Contents for 85106D
Page 11: ...x HP 85106D mm Wave System ...
Page 117: ...4 24 HP 85106D mm Wave System Service and Troubleshooting Check Unratioed Power Levels ...
Page 139: ...5 22 HP 85106D mm Wave System HP 85104A Test Set Module Troubleshooting Procedures ...
Page 140: ......
Page 141: ......
Page 142: ......
Page 143: ......
Page 208: ......
Page 238: ...B 2 HP 85106D mm Wave System Upgrade Paths ...
Page 243: ...HP 85106D mm Wave System v work surface installation 1 12 ...
Page 244: ... vi HP 85106D mm Wave System ...