5-18
HP 85106D mm-Wave System
HP 85104A Test Set Module
Troubleshooting Procedures
Figure 5-6 Power Level Test Setup
NOTE
It may be necessary to connect a waveguide straight section to the end of the
power sensor to reach the A13 output.
6. Reconnect the cables from the HP 85104A to the HP 85105A. Turn the
HP 85105A ON. Read the power level on the power meter. Refer to the
Table 5-3 on page 5-16 for the correct power level of the band you are
using.
7. Reverse the steps in this procedure to reassemble the test set module.
8. Return to the troubleshooting flowchart.
Table 5-4
RF Power Out of the A13 Source Block
Band (GHz)
Power Level (dBm)
33 to 50 (Q-band)
>3
40 to 60 (U-band)
>3
50 to 75 (V-band)
>3
75 to 110 (W-band)
>0