HP 85106D mm-Wave System
2-35
Operation
Common Problems and Their Solutions
Problem
System locks up soon after loading a machine dump.
Possible Causes and Solutions
If the system is configured and is using a synthesizer for the LO and a
machine dump expecting a sweeper LO is loaded, the system may hang
while attempting to phase lock. To recover from this situation:
1. Turn all instruments off.
2. While holding down a key on the HP 8510 numerical keypad, turn the
HP 8510 back on.
3. The HP 8510 starts the error diagnostic routine. Press [
=MARKER
].
4. Run the main program by pressing [
5
] [
1
].
Once the HP 8510 is sweeping, reset the phase lock by:
a.
Press [
SYSTEM
] {
MORE
} {
SYSTEM PHASELOCK
} {
NONE
}.
b.
Turn on the other instruments and reload the machine dump using
the correct MD_SYNWRxx file.