164
4
Maintaining the MSD
To reinstall the heater and sensor in the ion source
To reinstall the heater and sensor in the ion source
Materials needed:
Gloves, clean, lint-free
large (8650-0030)
small (8650-0029)
Hex ball driver, 1.5-mm (8710-1570)
Hex ball driver, 2.0-mm (8710-1804)
Hex nut driver, 5.5-mm (8710-1220)
Source heater assembly (G1099-60177)
1
Unpack the new source heater assembly.
The heater, temperature sensor, and heater block are already assembled.
2
Reinstall the repeller, repeller insulators, washer, and repeller nut.
The resulting assembly is called the repeller assembly.
3
Connect the repeller assembly to the source body.
4
Reinstall the filaments.
5
Reinstall the ion source in the source radiator (page 156).
Do not forget to reconnect the wires from the feedthrough board to the ion source.
Do not forget to reconnect the heater and temperature sensor wires to the
feedthrough board.
6
Close the vacuum manifold (page 56).
7
Pump down the MSD (page 58).
& $ 8 7 , 2 1
Do not overtighten the repeller nut, or the ceramic repeller insulators will break when the
source heats up. The nut should only be “finger-tight”.
Summary of Contents for 5973
Page 1: ...H Hardware Manual HP 5973 Mass Selective Detector ...
Page 3: ...HP 5973 Mass Selective Detector Hardware Manual ...
Page 25: ...1 Installing GC Columns To condition a capillary column 25 ...
Page 30: ...30 ...
Page 37: ...2 Operating the MSD To view MSD analyzer temperature and vacuum status 37 ...
Page 39: ...2 Operating the MSD To set monitors for MSD temperature and vacuum status 39 ...
Page 41: ...2 Operating the MSD To set the MSD analyzer temperatures 41 ...
Page 43: ...2 Operating the MSD To set the interface temperature from the ChemStation 43 ...
Page 51: ...2 Operating the MSD To remove the MSD covers 51 Upper cover Lower cover ...
Page 92: ...92 ...
Page 105: ...4 Maintaining the MSD To drain the foreline pump 105 Fill cap Drain plug Pump motor ...
Page 113: ...4 Maintaining the MSD To separate the MSD from the GC 113 ...
Page 121: ...4 Maintaining the MSD To reconnect the MSD to the GC 121 ...
Page 141: ...4 Maintaining the MSD To lubricate the side plate O ring 141 Side plate O ring ...
Page 159: ...4 Maintaining the MSD To remove a filament 159 Filament 2 Filament 1 ...
Page 161: ...4 Maintaining the MSD To reinstall a filament 161 Filament 2 Filament 1 ...
Page 210: ...210 ...
Page 214: ...214 ...
Page 232: ...232 7 Analyzer Analyzer heaters and radiators ...
Page 280: ...280 ...
Page 294: ...294 Index ...
Page 296: ...294 Chapter ...
Page 297: ...H Manual Part Number G1099 90027 Copyright 1998 Hewlett Packard Printed in USA 11 98 ...