56
2
Operating the MSD
To close the vacuum manifold
To close the vacuum manifold
Materials needed:
Gloves, clean, lint-free
large (8650-0030)
small (8650-0029)
1
Make sure all the internal analyzer electrical leads are correctly attached.
2
Check the side plate O-ring.
Make sure the O-ring has a
very
light coat of Apiezon high vacuum grease. If the
O-ring is very dry, it may not seal well. If the O-ring looks shiny, it has too much
grease on it. See page 140 for instructions for lubricating the side plate O-ring.
3
Close the side plate.
4
Reconnect the side board control cable and source power cable to the side
board.
5
Make sure the vent valve is closed.
6
Pump down the MSD (page 58).
7
Gently
hand tighten the front side plate thumbscrew
This is only necessary for CI MSDs, or if hydrogen or other flammable or toxic sub-
stance is used for carrier gas.
8
Once the MSD has pumped down, reinstall the upper MSD cover.
Wait until after pumpdown to reinstall the upper MSD cover.
: $ 5 1 , 1 *
7KLVWKXPEVFUHZPXVWEHIDVWHQHGIRU&,RSHUDWLRQRULIK\GURJHQRURWKHUKD]DUGRXVJDV
LVEHLQJXVHGDVWKH*&FDUULHUJDV,QWKHXQOLNHO\HYHQWRIDQH[SORVLRQLWPD\SUHYHQWWKH
VLGHSODWHIURPRSHQLQJ
& $ 8 7 , 2 1
Do not overtighten the thumbscrew; it can cause air leaks or prevent successful
pumpdown. Do not use a screwdriver to tighten the thumbscrew.
Summary of Contents for 5973
Page 1: ...H Hardware Manual HP 5973 Mass Selective Detector ...
Page 3: ...HP 5973 Mass Selective Detector Hardware Manual ...
Page 25: ...1 Installing GC Columns To condition a capillary column 25 ...
Page 30: ...30 ...
Page 37: ...2 Operating the MSD To view MSD analyzer temperature and vacuum status 37 ...
Page 39: ...2 Operating the MSD To set monitors for MSD temperature and vacuum status 39 ...
Page 41: ...2 Operating the MSD To set the MSD analyzer temperatures 41 ...
Page 43: ...2 Operating the MSD To set the interface temperature from the ChemStation 43 ...
Page 51: ...2 Operating the MSD To remove the MSD covers 51 Upper cover Lower cover ...
Page 92: ...92 ...
Page 105: ...4 Maintaining the MSD To drain the foreline pump 105 Fill cap Drain plug Pump motor ...
Page 113: ...4 Maintaining the MSD To separate the MSD from the GC 113 ...
Page 121: ...4 Maintaining the MSD To reconnect the MSD to the GC 121 ...
Page 141: ...4 Maintaining the MSD To lubricate the side plate O ring 141 Side plate O ring ...
Page 159: ...4 Maintaining the MSD To remove a filament 159 Filament 2 Filament 1 ...
Page 161: ...4 Maintaining the MSD To reinstall a filament 161 Filament 2 Filament 1 ...
Page 210: ...210 ...
Page 214: ...214 ...
Page 232: ...232 7 Analyzer Analyzer heaters and radiators ...
Page 280: ...280 ...
Page 294: ...294 Index ...
Page 296: ...294 Chapter ...
Page 297: ...H Manual Part Number G1099 90027 Copyright 1998 Hewlett Packard Printed in USA 11 98 ...