Installing GC Columns
To condition a capillary column
To condition a capillary column
Materials needed:
Carrier gas, (99.999% pure or better)
Wrench, open-end, 1/4-inch
×
5/16-inch (8710-0510)
1
Install the column in the GC inlet, page 22.
2
Allow the carrier gas to flow through the column for 5 minutes without
heating GC oven.
3
Ramp the oven temperature at 5°C/minute to 10°C above your highest
analytical temperature.
4
Hold this temperature. Allow the carrier gas to flow for several hours.
5
Return the GC oven temperature to a low standby temperature.
See Also
For more information about installing a capillary column, refer to the
Hewlett-Packard application note:
Optimizing splitless injections on your GC for high performance MS analysis
,
publication number 5091-4587E.
: $ 5 1 , 1 *
'RQRWFRQGLWLRQ\RXUFDSLOODU\FROXPQZLWKK\\GURJHQDFFXPXODWLRQLQWKH*&
RYHQFDQUHVXOWLQDQH[SORVLRQ,I\RXSODQWRXVHK\GURJHQDV\RXUFDUULHUJDVILUVW
FRQGLWLRQWKHFROXPQZLWKXOWUDSXUHRUEHWWHULQHUWJDVVXFKDVKHOLXPQLWURJHQ
RUDUJRQ
& $ 8 7 , 2 1
Do not exceed the maximum temperature rating of the column.
Summary of Contents for 5973
Page 1: ...H Hardware Manual HP 5973 Mass Selective Detector ...
Page 3: ...HP 5973 Mass Selective Detector Hardware Manual ...
Page 25: ...1 Installing GC Columns To condition a capillary column 25 ...
Page 30: ...30 ...
Page 37: ...2 Operating the MSD To view MSD analyzer temperature and vacuum status 37 ...
Page 39: ...2 Operating the MSD To set monitors for MSD temperature and vacuum status 39 ...
Page 41: ...2 Operating the MSD To set the MSD analyzer temperatures 41 ...
Page 43: ...2 Operating the MSD To set the interface temperature from the ChemStation 43 ...
Page 51: ...2 Operating the MSD To remove the MSD covers 51 Upper cover Lower cover ...
Page 92: ...92 ...
Page 105: ...4 Maintaining the MSD To drain the foreline pump 105 Fill cap Drain plug Pump motor ...
Page 113: ...4 Maintaining the MSD To separate the MSD from the GC 113 ...
Page 121: ...4 Maintaining the MSD To reconnect the MSD to the GC 121 ...
Page 141: ...4 Maintaining the MSD To lubricate the side plate O ring 141 Side plate O ring ...
Page 159: ...4 Maintaining the MSD To remove a filament 159 Filament 2 Filament 1 ...
Page 161: ...4 Maintaining the MSD To reinstall a filament 161 Filament 2 Filament 1 ...
Page 210: ...210 ...
Page 214: ...214 ...
Page 232: ...232 7 Analyzer Analyzer heaters and radiators ...
Page 280: ...280 ...
Page 294: ...294 Index ...
Page 296: ...294 Chapter ...
Page 297: ...H Manual Part Number G1099 90027 Copyright 1998 Hewlett Packard Printed in USA 11 98 ...