Troubleshooting the MSD
Chromatographic symptoms
Chromatographic symptoms
This section describes symptoms you may observe in the chromatograms gener-
ated by data acquisition. In general, these symptoms do not prevent you from
operating your GC/MSD system. They indicate, however, that the data you are
acquiring may not be the best data obtainable. These symptoms can be caused by
instrument malfunctions but are more likely caused by incorrect chromatographic
technique.
Two of the symptoms:
If sensitivity is low
and
If repeatability is poor
, also apply
to mass spectral data.
No peaks
If an analysis shows no chromatographic peaks, only a flat baseline or minor noise,
run one of the automated tune programs. If the MSD passes tune, the problem is
most likely related to the GC. If the MSD does not pass tune, the problem is most
likely in the MSD.
Passes tune
•
Incorrect sample concentration
•
No analytes present
•
Syringe missing from the ALS or not installed correctly
•
Injection accidentally made in split mode instead of splitless mode
•
Empty or almost empty sample vial
•
Dirty injection port
•
Leaking injection port*
•
Loose column nut at the injection port*
* These could cause a fault condition in the GC that would prevent the GC from
operating.
Summary of Contents for 5973
Page 1: ...H Hardware Manual HP 5973 Mass Selective Detector ...
Page 3: ...HP 5973 Mass Selective Detector Hardware Manual ...
Page 25: ...1 Installing GC Columns To condition a capillary column 25 ...
Page 30: ...30 ...
Page 37: ...2 Operating the MSD To view MSD analyzer temperature and vacuum status 37 ...
Page 39: ...2 Operating the MSD To set monitors for MSD temperature and vacuum status 39 ...
Page 41: ...2 Operating the MSD To set the MSD analyzer temperatures 41 ...
Page 43: ...2 Operating the MSD To set the interface temperature from the ChemStation 43 ...
Page 51: ...2 Operating the MSD To remove the MSD covers 51 Upper cover Lower cover ...
Page 92: ...92 ...
Page 105: ...4 Maintaining the MSD To drain the foreline pump 105 Fill cap Drain plug Pump motor ...
Page 113: ...4 Maintaining the MSD To separate the MSD from the GC 113 ...
Page 121: ...4 Maintaining the MSD To reconnect the MSD to the GC 121 ...
Page 141: ...4 Maintaining the MSD To lubricate the side plate O ring 141 Side plate O ring ...
Page 159: ...4 Maintaining the MSD To remove a filament 159 Filament 2 Filament 1 ...
Page 161: ...4 Maintaining the MSD To reinstall a filament 161 Filament 2 Filament 1 ...
Page 210: ...210 ...
Page 214: ...214 ...
Page 232: ...232 7 Analyzer Analyzer heaters and radiators ...
Page 280: ...280 ...
Page 294: ...294 Index ...
Page 296: ...294 Chapter ...
Page 297: ...H Manual Part Number G1099 90027 Copyright 1998 Hewlett Packard Printed in USA 11 98 ...