GC/MSD interface
The GC/MSD interface is a heated conduit into the MSD for the capillary
column. It is bolted onto the right side of the vacuum manifold, with an
O-ring seal. A channel machined into the flange for the seal provides thermal
isolation between the heated interface and the O-ring and manifold. The GC/
MSD interface is covered by a protective cover which should be left in place.
One end of the GC/MSD interface passes through the side of the gas
chromatograph and extends into the GC oven. This end is threaded
(thread size is 10
×
32), allowing connection of the column with a nut and
ferrule. The other end of the GC/MSD interface fits into the ion source. The
last two millimeters of the capillary column extend past the end of the guide
tube and into the ionization chamber.
The GC/MSD interface is heated by an electric cartridge heater. The heater
is powered and controlled by Thermal Aux #2 heated zone of the HP 6890
Series GC. The GC/MSD interface temperature can be set from the MSD
ChemStation or from the keypad of the gas chromatograph. A sensor (ther-
mocouple) in the GC/MSD interface monitors the temperature.
The GC/MSD interface should be operated in the 250
°
to 350
°
C range.
Subject to that restriction, the GC/MSD interface temperature should be
slightly higher than the maximum GC oven temperature, but
never
higher
than the maximum column temperature.
See Also
To install a capillary column using the installation tool,
To install a capillary column without the installation tool,
: $ 5 1 , 1 *
7KH*&06'LQWHUIDFHRSHUDWHVDWKLJKWHPSHUDWXUHV,I\RXWRXFKLWZKHQLWLVKRWLWZLOO
EXUQ\RX
Summary of Contents for 5973
Page 1: ...H Hardware Manual HP 5973 Mass Selective Detector ...
Page 3: ...HP 5973 Mass Selective Detector Hardware Manual ...
Page 25: ...1 Installing GC Columns To condition a capillary column 25 ...
Page 30: ...30 ...
Page 37: ...2 Operating the MSD To view MSD analyzer temperature and vacuum status 37 ...
Page 39: ...2 Operating the MSD To set monitors for MSD temperature and vacuum status 39 ...
Page 41: ...2 Operating the MSD To set the MSD analyzer temperatures 41 ...
Page 43: ...2 Operating the MSD To set the interface temperature from the ChemStation 43 ...
Page 51: ...2 Operating the MSD To remove the MSD covers 51 Upper cover Lower cover ...
Page 92: ...92 ...
Page 105: ...4 Maintaining the MSD To drain the foreline pump 105 Fill cap Drain plug Pump motor ...
Page 113: ...4 Maintaining the MSD To separate the MSD from the GC 113 ...
Page 121: ...4 Maintaining the MSD To reconnect the MSD to the GC 121 ...
Page 141: ...4 Maintaining the MSD To lubricate the side plate O ring 141 Side plate O ring ...
Page 159: ...4 Maintaining the MSD To remove a filament 159 Filament 2 Filament 1 ...
Page 161: ...4 Maintaining the MSD To reinstall a filament 161 Filament 2 Filament 1 ...
Page 210: ...210 ...
Page 214: ...214 ...
Page 232: ...232 7 Analyzer Analyzer heaters and radiators ...
Page 280: ...280 ...
Page 294: ...294 Index ...
Page 296: ...294 Chapter ...
Page 297: ...H Manual Part Number G1099 90027 Copyright 1998 Hewlett Packard Printed in USA 11 98 ...