Installing GC Columns
To install a capillary column using the installation tool
To install a capillary column using the installation tool
Materials needed:
Column cutter (5181-8836)
Column installation tool (G1099-20030)
Ferrules
0.3-mm id, for 0.10-mm id columns (5062-3507)
0.4-mm id, for 0.20- and 0.25-mm id columns (5062-3508)
0.5-mm id, for 0.32-mm id columns (5062-3506)
0.8-mm id, for 0.53-mm id columns (5062-3538)
Gloves, clean
large (8650-0030)
small (8650-0029)
Interface column nut (05988-20066)
Septum (may be old, used inlet septum)
Wrenches, open-end, 1/4-inch
×
5/16-inch (8710-0510) – 2 required
For optimal sensitivity performance, see
To install a capillary column without
the installation tool
1
Slide a septum, interface
column nut, and conditioned ferrule onto the
free end of the column.
The tapered end of the ferrule should point toward the nut.
2
Insert the column into the column installation tool.
Slide the column through until the end extends past the end of the tool.
3
Cut 1 cm off the end of the column (page 20).
4
Position the column so that 1 to 2 mm extends past the end of the tool.
Hand tighten the nut.
5
Slide the septum to touch the end of the nut.
The septum will help assure that the position is correct.
6
Use two wrenches to tighten the nut 1/4 turn.
The column should not slide when tugged
gently
.
7
Remove the column and nut from the installation tool.
The total length from the septum to the end of the column is 176 mm.
Summary of Contents for 5973
Page 1: ...H Hardware Manual HP 5973 Mass Selective Detector ...
Page 3: ...HP 5973 Mass Selective Detector Hardware Manual ...
Page 25: ...1 Installing GC Columns To condition a capillary column 25 ...
Page 30: ...30 ...
Page 37: ...2 Operating the MSD To view MSD analyzer temperature and vacuum status 37 ...
Page 39: ...2 Operating the MSD To set monitors for MSD temperature and vacuum status 39 ...
Page 41: ...2 Operating the MSD To set the MSD analyzer temperatures 41 ...
Page 43: ...2 Operating the MSD To set the interface temperature from the ChemStation 43 ...
Page 51: ...2 Operating the MSD To remove the MSD covers 51 Upper cover Lower cover ...
Page 92: ...92 ...
Page 105: ...4 Maintaining the MSD To drain the foreline pump 105 Fill cap Drain plug Pump motor ...
Page 113: ...4 Maintaining the MSD To separate the MSD from the GC 113 ...
Page 121: ...4 Maintaining the MSD To reconnect the MSD to the GC 121 ...
Page 141: ...4 Maintaining the MSD To lubricate the side plate O ring 141 Side plate O ring ...
Page 159: ...4 Maintaining the MSD To remove a filament 159 Filament 2 Filament 1 ...
Page 161: ...4 Maintaining the MSD To reinstall a filament 161 Filament 2 Filament 1 ...
Page 210: ...210 ...
Page 214: ...214 ...
Page 232: ...232 7 Analyzer Analyzer heaters and radiators ...
Page 280: ...280 ...
Page 294: ...294 Index ...
Page 296: ...294 Chapter ...
Page 297: ...H Manual Part Number G1099 90027 Copyright 1998 Hewlett Packard Printed in USA 11 98 ...