Troubleshooting the MSD
Pressure symptoms
Foreline pressure is too low
If the pressures you observe are below 20 mTorr, check for the following:
•
Column (carrier gas) flow is too low
•
Column plugged or crushed by an overtightened nut
•
Empty or insufficient carrier gas supply*
•
Bent or pinched carrier gas tubing*
•
Foreline gauge is not working correctly
* These could create a fault condition in the GC that would prevent the GC from
operating.
Vacuum manifold pressure is too low
If the pressures you observe are below 1
×
10
-6
Torr, check for the following:
•
Column (carrier gas) flow is too low
•
Column plugged or crushed by overtightened nut
•
Empty or insufficient carrier gas supply*
•
Bent or pinched carrier gas tubing*
* These could create a fault condition in the GC that would prevent the GC from
operating.
Gauge controller displays 9.9+9 and then goes blank
This indicates the pressure in the vacuum manifold is above 8
×
10
-3
Torr.
•
Solvent peak from an on-column injection
•
MSD has not had enough time to pump down
•
Excessive foreline pressure
•
Bad connection or bad cable between the triode gauge and gauge controller
•
Diffusion pump fluid level is low or fluid is contaminated
•
Triode gauge tube has failed
•
Line voltage too low
•
Turbomolecular pump is not working correctly
Summary of Contents for 5973
Page 1: ...H Hardware Manual HP 5973 Mass Selective Detector ...
Page 3: ...HP 5973 Mass Selective Detector Hardware Manual ...
Page 25: ...1 Installing GC Columns To condition a capillary column 25 ...
Page 30: ...30 ...
Page 37: ...2 Operating the MSD To view MSD analyzer temperature and vacuum status 37 ...
Page 39: ...2 Operating the MSD To set monitors for MSD temperature and vacuum status 39 ...
Page 41: ...2 Operating the MSD To set the MSD analyzer temperatures 41 ...
Page 43: ...2 Operating the MSD To set the interface temperature from the ChemStation 43 ...
Page 51: ...2 Operating the MSD To remove the MSD covers 51 Upper cover Lower cover ...
Page 92: ...92 ...
Page 105: ...4 Maintaining the MSD To drain the foreline pump 105 Fill cap Drain plug Pump motor ...
Page 113: ...4 Maintaining the MSD To separate the MSD from the GC 113 ...
Page 121: ...4 Maintaining the MSD To reconnect the MSD to the GC 121 ...
Page 141: ...4 Maintaining the MSD To lubricate the side plate O ring 141 Side plate O ring ...
Page 159: ...4 Maintaining the MSD To remove a filament 159 Filament 2 Filament 1 ...
Page 161: ...4 Maintaining the MSD To reinstall a filament 161 Filament 2 Filament 1 ...
Page 210: ...210 ...
Page 214: ...214 ...
Page 232: ...232 7 Analyzer Analyzer heaters and radiators ...
Page 280: ...280 ...
Page 294: ...294 Index ...
Page 296: ...294 Chapter ...
Page 297: ...H Manual Part Number G1099 90027 Copyright 1998 Hewlett Packard Printed in USA 11 98 ...