Troubleshooting the MSD
Error messages
The system is in vent state
Although the message says the system is in vent state, if the fault has just occurred
the MSD may actually still be under vacuum and the diffusion pump may still be
hot or the turbo pump may still be at high speed. Wait at least 30 minutes after
seeing this message before you actually vent the MSD.
•
System was vented on purpose (no problem)
•
Fan fault has turned off the high vacuum pump (power cycle the MSD to clear
the fault)
•
Fuse for the high vacuum pump has failed
•
Diffusion pump heater cartridge has failed
•
MSD electronics are not working correctly
There is no emission current
•
Filament is not connected properly
•
Filament has failed
•
MSD electronics are not working correctly
There is not enough signal to begin tune
•
Poor mass axis calibration
•
Amu gain or offset is too high
•
Calibration vial is empty or almost empty
•
Excessive pressure in the vacuum manifold
•
Air leak
•
Electron multiplier voltage is too low
•
Signal cable is not connected
•
Electrical leads to the detector are not connected correctly
•
Electrical leads to the ion source are not connected correctly
•
Filament shorted to the source body
& $ 8 7 , 2 1
Venting the MSD too soon after this message appears can result in the backstreaming of
a diffusion pump or damage to a turbomolecular pump.
Summary of Contents for 5973
Page 1: ...H Hardware Manual HP 5973 Mass Selective Detector ...
Page 3: ...HP 5973 Mass Selective Detector Hardware Manual ...
Page 25: ...1 Installing GC Columns To condition a capillary column 25 ...
Page 30: ...30 ...
Page 37: ...2 Operating the MSD To view MSD analyzer temperature and vacuum status 37 ...
Page 39: ...2 Operating the MSD To set monitors for MSD temperature and vacuum status 39 ...
Page 41: ...2 Operating the MSD To set the MSD analyzer temperatures 41 ...
Page 43: ...2 Operating the MSD To set the interface temperature from the ChemStation 43 ...
Page 51: ...2 Operating the MSD To remove the MSD covers 51 Upper cover Lower cover ...
Page 92: ...92 ...
Page 105: ...4 Maintaining the MSD To drain the foreline pump 105 Fill cap Drain plug Pump motor ...
Page 113: ...4 Maintaining the MSD To separate the MSD from the GC 113 ...
Page 121: ...4 Maintaining the MSD To reconnect the MSD to the GC 121 ...
Page 141: ...4 Maintaining the MSD To lubricate the side plate O ring 141 Side plate O ring ...
Page 159: ...4 Maintaining the MSD To remove a filament 159 Filament 2 Filament 1 ...
Page 161: ...4 Maintaining the MSD To reinstall a filament 161 Filament 2 Filament 1 ...
Page 210: ...210 ...
Page 214: ...214 ...
Page 232: ...232 7 Analyzer Analyzer heaters and radiators ...
Page 280: ...280 ...
Page 294: ...294 Index ...
Page 296: ...294 Chapter ...
Page 297: ...H Manual Part Number G1099 90027 Copyright 1998 Hewlett Packard Printed in USA 11 98 ...