132
4
Maintaining the MSD
To reinstall a calibration valve
To reinstall a calibration valve
Materials needed:
Calibration valve
for diffusion pump (G1099-60200)
for turbomolecular pump (G1099-60203)
O-ring, for calibration valve (0905-1217) – replace if the old O-ring is damaged
PFTBA (05971-60571) or other tuning compound
Screwdriver, TORX T-15 (8710-1622)
1
Remove the old calibration valve (page 130).
2
Make sure the calibration valve O-ring is in place.
If the O-ring is worn or damaged, replace it.
3
Install the calibration valve.
Tighten the screws that hold it in place. Make sure you use the calibration valve
that matches the high vacuum pump in your MSD. The different calibration valves
have different restrictors. Using the wrong valve will interfere with tuning.
4
Reconnect the calibration valve cable to the connector next to the fan.
5
Remove the vial from the new calibration valve (page 122).
The valve is supplied with a vial already installed.
6
Fill and reinstall the calibration vial (page 124).
7
Pump down the MSD (page 58).
8
Select
Purge Cal Valve
from the Vacuum menu in the Diagnostics/Vacuum
Control view.
& $ 8 7 , 2 1
Failure to purge the calibration valve will result in damage to the filaments and detector.
Summary of Contents for 5973
Page 1: ...H Hardware Manual HP 5973 Mass Selective Detector ...
Page 3: ...HP 5973 Mass Selective Detector Hardware Manual ...
Page 25: ...1 Installing GC Columns To condition a capillary column 25 ...
Page 30: ...30 ...
Page 37: ...2 Operating the MSD To view MSD analyzer temperature and vacuum status 37 ...
Page 39: ...2 Operating the MSD To set monitors for MSD temperature and vacuum status 39 ...
Page 41: ...2 Operating the MSD To set the MSD analyzer temperatures 41 ...
Page 43: ...2 Operating the MSD To set the interface temperature from the ChemStation 43 ...
Page 51: ...2 Operating the MSD To remove the MSD covers 51 Upper cover Lower cover ...
Page 92: ...92 ...
Page 105: ...4 Maintaining the MSD To drain the foreline pump 105 Fill cap Drain plug Pump motor ...
Page 113: ...4 Maintaining the MSD To separate the MSD from the GC 113 ...
Page 121: ...4 Maintaining the MSD To reconnect the MSD to the GC 121 ...
Page 141: ...4 Maintaining the MSD To lubricate the side plate O ring 141 Side plate O ring ...
Page 159: ...4 Maintaining the MSD To remove a filament 159 Filament 2 Filament 1 ...
Page 161: ...4 Maintaining the MSD To reinstall a filament 161 Filament 2 Filament 1 ...
Page 210: ...210 ...
Page 214: ...214 ...
Page 232: ...232 7 Analyzer Analyzer heaters and radiators ...
Page 280: ...280 ...
Page 294: ...294 Index ...
Page 296: ...294 Chapter ...
Page 297: ...H Manual Part Number G1099 90027 Copyright 1998 Hewlett Packard Printed in USA 11 98 ...