124
4
Maintaining the MSD
To refill and reinstall the calibration vial
To refill and reinstall the calibration vial
Materials needed:
PFTBA (05971-60571) or other tuning compound
1
Remove the calibration vial (page 122).
2
Pour PFTBA into the vial.
Leave 6-mm of the vial unfilled.
3
Push the calibration vial into the valve as far as possible.
4
Withdraw the vial 1 mm.
This prevents damage when you tighten the collar.
5
Turn the collar clockwise to tighten it.
Clockwise as viewed from the bottom (vial side) of the collar. The collar should be
snug but not overly tight. Do
not
use a tool to tighten the collar. It does not
require that much force.
6
Reinstall the upper MSD cover.
7
Select
Purge Cal Valve
from the Vacuum menu in the Diagnostics/Vacuum
Control view.
& $ 8 7 , 2 1
Failure to purge the calibration valve will result in damage to the filaments and detector.
Summary of Contents for 5973
Page 1: ...H Hardware Manual HP 5973 Mass Selective Detector ...
Page 3: ...HP 5973 Mass Selective Detector Hardware Manual ...
Page 25: ...1 Installing GC Columns To condition a capillary column 25 ...
Page 30: ...30 ...
Page 37: ...2 Operating the MSD To view MSD analyzer temperature and vacuum status 37 ...
Page 39: ...2 Operating the MSD To set monitors for MSD temperature and vacuum status 39 ...
Page 41: ...2 Operating the MSD To set the MSD analyzer temperatures 41 ...
Page 43: ...2 Operating the MSD To set the interface temperature from the ChemStation 43 ...
Page 51: ...2 Operating the MSD To remove the MSD covers 51 Upper cover Lower cover ...
Page 92: ...92 ...
Page 105: ...4 Maintaining the MSD To drain the foreline pump 105 Fill cap Drain plug Pump motor ...
Page 113: ...4 Maintaining the MSD To separate the MSD from the GC 113 ...
Page 121: ...4 Maintaining the MSD To reconnect the MSD to the GC 121 ...
Page 141: ...4 Maintaining the MSD To lubricate the side plate O ring 141 Side plate O ring ...
Page 159: ...4 Maintaining the MSD To remove a filament 159 Filament 2 Filament 1 ...
Page 161: ...4 Maintaining the MSD To reinstall a filament 161 Filament 2 Filament 1 ...
Page 210: ...210 ...
Page 214: ...214 ...
Page 232: ...232 7 Analyzer Analyzer heaters and radiators ...
Page 280: ...280 ...
Page 294: ...294 Index ...
Page 296: ...294 Chapter ...
Page 297: ...H Manual Part Number G1099 90027 Copyright 1998 Hewlett Packard Printed in USA 11 98 ...