Troubleshooting the MSD
Contamination
Contamination
Contamination is usually identified by excessive background in the mass spectra.
It can come from the GC or from the MSD. The source of the contamination can
sometimes be determined by identifying the contaminants. Some contaminants
are much more likely to originate in the GC. Others are more likely to originate in
the MSD.
Contamination originating in the GC typically comes from one of these sources:
•
Column or septum bleed
•
Dirty injection port
•
Injection port liner
•
Contaminated syringe
•
Poor quality carrier gas
•
Dirty carrier gas tubing
•
Fingerprints (improper handling of clean parts)
Contamination originating in the MSD typically comes from one of the following
sources:
•
Air leak
•
Cleaning solvents and materials
•
Diffusion pump fluid
•
Foreline pump oil
•
Fingerprints (improper handling of clean parts)
The following table lists some of the more common contaminants, the ions charac-
teristic of those contaminants, and the likely sources of those contaminants.
Summary of Contents for 5973
Page 1: ...H Hardware Manual HP 5973 Mass Selective Detector ...
Page 3: ...HP 5973 Mass Selective Detector Hardware Manual ...
Page 25: ...1 Installing GC Columns To condition a capillary column 25 ...
Page 30: ...30 ...
Page 37: ...2 Operating the MSD To view MSD analyzer temperature and vacuum status 37 ...
Page 39: ...2 Operating the MSD To set monitors for MSD temperature and vacuum status 39 ...
Page 41: ...2 Operating the MSD To set the MSD analyzer temperatures 41 ...
Page 43: ...2 Operating the MSD To set the interface temperature from the ChemStation 43 ...
Page 51: ...2 Operating the MSD To remove the MSD covers 51 Upper cover Lower cover ...
Page 92: ...92 ...
Page 105: ...4 Maintaining the MSD To drain the foreline pump 105 Fill cap Drain plug Pump motor ...
Page 113: ...4 Maintaining the MSD To separate the MSD from the GC 113 ...
Page 121: ...4 Maintaining the MSD To reconnect the MSD to the GC 121 ...
Page 141: ...4 Maintaining the MSD To lubricate the side plate O ring 141 Side plate O ring ...
Page 159: ...4 Maintaining the MSD To remove a filament 159 Filament 2 Filament 1 ...
Page 161: ...4 Maintaining the MSD To reinstall a filament 161 Filament 2 Filament 1 ...
Page 210: ...210 ...
Page 214: ...214 ...
Page 232: ...232 7 Analyzer Analyzer heaters and radiators ...
Page 280: ...280 ...
Page 294: ...294 Index ...
Page 296: ...294 Chapter ...
Page 297: ...H Manual Part Number G1099 90027 Copyright 1998 Hewlett Packard Printed in USA 11 98 ...